Greater accuracy in positioning of the acetabular cup by using an image-free navigation system
In a prospective and randomised clinical study, acetabular cups were implanted free-hand (control group n=22) or with computer assistance using an image-free navigation system (study group n=23). The cup position was determined postoperatively on pelvic CT. An average inclination of 42.3 degrees (ra...
Gespeichert in:
Veröffentlicht in: | International orthopaedics 2005-10, Vol.29 (5), p.272-276 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | In a prospective and randomised clinical study, acetabular cups were implanted free-hand (control group n=22) or with computer assistance using an image-free navigation system (study group n=23). The cup position was determined postoperatively on pelvic CT. An average inclination of 42.3 degrees (range: 30 degrees -53 degrees ; SD+/-7.0 degrees ) and an average anteversion of 24.0 degrees (range: -3 degrees to 51 degrees ; SD+/-15.0 degrees ) were found in the control group, and an average inclination of 45.0 degrees (range: 40 degrees -50 degrees ; SD+/-2.8 degrees ) and an average anteversion of 14.4 degrees (range: 5 degrees -25 degrees ; SS+/-5.0 degrees ) in the computer-assisted study group. The deviations from the desired cup position (45 degrees inclination, 15 degrees anteversion) were significantly lower in the computer-assisted study group (p |
---|---|