Greater accuracy in positioning of the acetabular cup by using an image-free navigation system

In a prospective and randomised clinical study, acetabular cups were implanted free-hand (control group n=22) or with computer assistance using an image-free navigation system (study group n=23). The cup position was determined postoperatively on pelvic CT. An average inclination of 42.3 degrees (ra...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:International orthopaedics 2005-10, Vol.29 (5), p.272-276
Hauptverfasser: Kalteis, T, Handel, M, Herold, T, Perlick, L, Baethis, H, Grifka, J
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:In a prospective and randomised clinical study, acetabular cups were implanted free-hand (control group n=22) or with computer assistance using an image-free navigation system (study group n=23). The cup position was determined postoperatively on pelvic CT. An average inclination of 42.3 degrees (range: 30 degrees -53 degrees ; SD+/-7.0 degrees ) and an average anteversion of 24.0 degrees (range: -3 degrees to 51 degrees ; SD+/-15.0 degrees ) were found in the control group, and an average inclination of 45.0 degrees (range: 40 degrees -50 degrees ; SD+/-2.8 degrees ) and an average anteversion of 14.4 degrees (range: 5 degrees -25 degrees ; SS+/-5.0 degrees ) in the computer-assisted study group. The deviations from the desired cup position (45 degrees inclination, 15 degrees anteversion) were significantly lower in the computer-assisted study group (p