Inadvertent Port: Catheter Placement in Azygos Vein

Abstract Totally implantable ports are safe and effective means of venous access for administration of chemotherapy. One of the usual vessels accessed, through which the port is placed, is the subclavian. Herein, we report a case where the central access was obtained through the left subclavian vein...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The International journal of angiology 2012-06, Vol.21 (2), p.103-106
Hauptverfasser: Harish, K., Madhu, Y. C.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Abstract Totally implantable ports are safe and effective means of venous access for administration of chemotherapy. One of the usual vessels accessed, through which the port is placed, is the subclavian. Herein, we report a case where the central access was obtained through the left subclavian vein. But the catheter when it was placed was found to be in the azygos vein. This was confirmed with dye study. This is the first report of such an occurrence with subclavian access. The catheter was later withdrawn and repositioned in the superior vena cava. We recommend that the entire procedure of catheter placement must be done under fluoroscopy guidance to ensure safe and error-free positioning.
ISSN:1061-1711
1615-5939
DOI:10.1055/s-0032-1315628