Initial Experience with Intracranial Stent-Graft Use: Technical Notes

We describe our initial experience with the placement of two premounted balloon expandable intracranial Jostent stent-grafts within the intracavernous internal carotid artery for the treatment of a symptomatic large intracavernous aneurysm in one case and a post-traumatic caroticocavernous fistula i...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Interventional neuroradiology 2005-06, Vol.11 (2), p.131-139
Hauptverfasser: Duncan, I. C., Fourie, P. A.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We describe our initial experience with the placement of two premounted balloon expandable intracranial Jostent stent-grafts within the intracavernous internal carotid artery for the treatment of a symptomatic large intracavernous aneurysm in one case and a post-traumatic caroticocavernous fistula in the second. Among the initial technical complications we encountered were stent-graft migration and rapidly progressive intragraft thrombosis, with delayed sealing of the stent-graft coverings and exclusion of the pathologies relating to the use of abciximab in both cases. Despite these initial problems both cases had excellent short-term clinical outcomes with angiographic exclusion of both lesions by day three and good clinical and angiographic outcomes at one and two months respectively.