Initial Experience with Intracranial Stent-Graft Use: Technical Notes
We describe our initial experience with the placement of two premounted balloon expandable intracranial Jostent stent-grafts within the intracavernous internal carotid artery for the treatment of a symptomatic large intracavernous aneurysm in one case and a post-traumatic caroticocavernous fistula i...
Gespeichert in:
Veröffentlicht in: | Interventional neuroradiology 2005-06, Vol.11 (2), p.131-139 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We describe our initial experience with the placement of two premounted balloon expandable intracranial Jostent stent-grafts within the intracavernous internal carotid artery for the treatment of a symptomatic large intracavernous aneurysm in one case and a post-traumatic caroticocavernous fistula in the second. Among the initial technical complications we encountered were stent-graft migration and rapidly progressive intragraft thrombosis, with delayed sealing of the stent-graft coverings and exclusion of the pathologies relating to the use of abciximab in both cases. Despite these initial problems both cases had excellent short-term clinical outcomes with angiographic exclusion of both lesions by day three and good clinical and angiographic outcomes at one and two months respectively. |
---|---|
ISSN: | 1591-0199 2385-2011 |
DOI: | 10.1177/159101990501100203 |