Fast scanning coaxial optoacoustic microscopy

The hybrid nature of optoacoustic imaging might impose limitations on concurrent placement of optical and ultrasonic detection components, especially in high resolution microscopic applications that require dense arrangements and miniaturization of components. This hinders optimal deployment of the...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Biomedical optics express 2012-07, Vol.3 (7), p.1724-1731
Hauptverfasser: Ma, Rui, Söntges, Sebastian, Shoham, Shy, Ntziachristos, Vasilis, Razansky, Daniel
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The hybrid nature of optoacoustic imaging might impose limitations on concurrent placement of optical and ultrasonic detection components, especially in high resolution microscopic applications that require dense arrangements and miniaturization of components. This hinders optimal deployment of the optical excitation and ultrasonic detection paths, leading to reduction of imaging speed and spatial resolution performance. We suggest a compact coaxial design for optoacoustic microscopy that allows optimizing both the light illumination and ultrasonic detection parameters of the imaging system. System performance is showcased in phantoms and in vivo imaging of microvasculature, achieving real time operation in two dimensions and penetration of 6 mm into optically dense human tissues.