Development and Advanced Validation of an Optimized Method for the Quantitation of Aβ42 in Human Cerebrospinal Fluid

Cerebrospinal fluid (CSF) biomarkers have been extensively utilized in the diagnosis of Alzheimer’s disease (AD) and characterization of progression. One important CSF biomarker is the amyloid beta 42 (Aβ 42 ) peptide, a key player in AD pathogenesis. The INNOTEST® Aβ 42 ELISA kit has been widely us...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The AAPS journal 2012-09, Vol.14 (3), p.510-518
Hauptverfasser: Cullen, Valerie C., Fredenburg, Ross A., Evans, Cindy, Conliffe, Phyllis R., Solomon, Michael E.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Cerebrospinal fluid (CSF) biomarkers have been extensively utilized in the diagnosis of Alzheimer’s disease (AD) and characterization of progression. One important CSF biomarker is the amyloid beta 42 (Aβ 42 ) peptide, a key player in AD pathogenesis. The INNOTEST® Aβ 42 ELISA kit has been widely used but an advanced level of method development and validation has not been reported. To support a clinical trial in AD, we successfully completed a Good Laboratory Practices (GLP)-level validation of the method to establish the parameters of precision, accuracy, parallelism, selectivity, specificity, and linearity of dilution of the assay in CSF matrix, as well as CSF storage stability. Several modifications were required to optimize the assay and ensure consistent results in a clinical-trial setting. These included the use of additional calibrators, an adjusted standard curve range, a minimum required dilution (MRD) of CSF by 6-fold to avoid matrix interference and mitigation of analyte adsorption to labware by the addition of Tween-20. The optimized method displayed a quantitative range of 375–4,500 pg/mL. The inter-assay precision was ≤12.1 % CV and the inter-assay relative accuracy was ≤10.9 % absolute bias, bringing the total error of the assay to ≤23 %. The intra-assay precision of the assay at the high validation standard and below was ≤5.5 % CV; this enables sensitive detection of biomarker changes across a therapeutic regime. The INNOTEST® Aβ 42 ELISA kit, modified as reported here, may be appropriate for many applications, including regulatory agency acceptable clinical diagnosis and pharmacodynamic assessment.
ISSN:1550-7416
1550-7416
DOI:10.1208/s12248-012-9360-7