Extended healing validation of an artificial tendon to connect the quadriceps muscle to the Tibia: 180-day study

Whenever a tendon or its bone insertion is disrupted or removed, existing surgical techniques provide a temporary connection or scaffolding to promote healing, but the interface of living to non‐living materials soon breaks down under the stress of these applications, if it must bear the load more t...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of orthopaedic research 2012-07, Vol.30 (7), p.1112-1117
Hauptverfasser: Melvin, Alan J., Litsky, Alan S., Mayerson, Joel L., Stringer, Keith, Juncosa-Melvin, Natalia
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Whenever a tendon or its bone insertion is disrupted or removed, existing surgical techniques provide a temporary connection or scaffolding to promote healing, but the interface of living to non‐living materials soon breaks down under the stress of these applications, if it must bear the load more than acutely. Patients are thus disabled whose prostheses, defect size, or mere anatomy limit the availability or outcomes of such treatments. Our group developed the OrthoCoupler™ device to join skeletal muscle to prosthetic or natural structures without this interface breakdown. In this study, the goat knee extensor mechanism (quadriceps tendon, patella, and patellar tendon) was removed from the right hind limb in 16 goats. The device connected the quadriceps muscle to a stainless steel bone plate on the tibia. Mechanical testing and histology specimens were collected from each operated leg and contralateral unoperated control legs at 180 days. Maximum forces in the operated leg (vs. unoperated) were 1,400 ± 93 N (vs. 1,179 ± 61 N), linear stiffnesses were 33 ± 3 N/mm (vs. 37 ± 4 N/mm), and elongations at failure were 92.1 ± 5.3 mm (vs. 68.4 ± 3.8 mm; mean ± SEM). Higher maximum forces (p = 0.02) and elongations at failure (p = 0.008) of legs with the device versus unoperated controls were significant; linear stiffnesses were not (p = 0.3). We believe this technology will yield improved procedures for clinical challenges in orthopedic oncology, revision arthroplasty, tendon transfer, and tendon injury reconstruction. © 2011 Orthopaedic Research Society. Published by Wiley Periodicals, Inc. J Orthop Res 30:1112–1117, 2012
ISSN:0736-0266