Quantification of irinotecan, SN38, and SN38G in human and porcine plasma by ultra high-performance liquid chromatography–tandem mass spectrometry and its application to hepatic chemoembolization

► We developed and validated a novel uHPLC–MS/MS assay for the simultaneous quantification of irinotecan, SN38, and SN38-G in both pig and human plasma. ► Clinically relevant calibration ranges for irinotecan (5–1000ng/mL), its metabolites (0.5–100ng/mL) were used. ► Successfully applied to both pre...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of pharmaceutical and biomedical analysis 2012-03, Vol.62, p.140-148
Hauptverfasser: Chen, Xiaohong, Peer, Cody J., Alfaro, Raul, Tian, Tian, Spencer, Shawn D., Figg, William D.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:► We developed and validated a novel uHPLC–MS/MS assay for the simultaneous quantification of irinotecan, SN38, and SN38-G in both pig and human plasma. ► Clinically relevant calibration ranges for irinotecan (5–1000ng/mL), its metabolites (0.5–100ng/mL) were used. ► Successfully applied to both preclinical and clinical studies undergoing hepatic chemoembolization. An analytical method was developed and validated for the quantitative determination of irinotecan, its active metabolite SN38, and glucuronidated SN38 (SN38-G) in both porcine and human plasma. Calibration curves were linear within the concentration range of 0.5–100ng/mL for SN38 and SN38-G, and 5–1000ng/mL for irinotecan. Sample pretreatment involved solid-phase extraction of 0.1mL aliquots of plasma. Irinotecan, SN38, SN38-G, and the internal standards, irinotecan-d10, tolbutamide, and camptothecin, respectively, were separated on a Waters ACQUITY UPLC™ BEH RP18 column (2.1mm×50mm, 1.7μm), using a mobile phase composed of methanol and 0.1% formic acid. Accuracy of quality control samples in human plasma ranged from 98.5 to 110.3%, 99.5 to 101.7% and 96.2 to 98.9% for irinotecan, SN38, and SN38-G, respectively. Precision of the three analytes in the same order ranged from 0.8 to 2.8%, 2.4 to 5.7%, and 2.4 to 2.8%. All three analytes proved stable in plasma through four freeze/thaw cycles, as well as through 6h in whole blood at room temperature. The method was likewise validated in porcine plasma with comparable accuracies and precisions also within the generally acceptable range. The validated method was applied to both preclinical and clinical trials involving hepatic chemoembolization of irinotecan drug-eluting beads to study the pharmacokinetics of the three analytes.
ISSN:0731-7085
1873-264X
1873-264X