Evoked Potential Monitoring Identifies Possible Neurological Injury During Positioning for Craniotomy
Somatosensory-evoked potential (SSEP) monitoring is commonly used to detect changes in nerve conduction and prevent impending nerve injury. We present a case series of two patients who had SSEP monitoring for their surgical craniotomy procedure, and who, upon positioning supine with their head tilte...
Gespeichert in:
Veröffentlicht in: | Anesthesia and analgesia 2009-09, Vol.109 (3), p.817-821 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Somatosensory-evoked potential (SSEP) monitoring is commonly used to detect changes in nerve conduction and prevent impending nerve injury. We present a case series of two patients who had SSEP monitoring for their surgical craniotomy procedure, and who, upon positioning supine with their head tilted 30°–45°, developed unilateral upper extremity SSEP changes. These SSEP changes were reversed when the patients were repositioned. These cases indicate the clinical usefulness of monitoring SSEPs while positioning the patient and adjusting position accordingly to prevent injury. |
---|---|
ISSN: | 0003-2999 1526-7598 |
DOI: | 10.1213/ane.0b013e3181b086bd |