A practical method of measuring the human temporal contrast sensitivity function
One of the more significant indicators of neural age-related loss and disease is reduced temporal processing speed. It would, therefore, be useful to have an accurate and practical device that measures the full range of an individual's temporal processing abilities (characterized as the tempora...
Gespeichert in:
Veröffentlicht in: | Biomedical optics express 2010-08, Vol.1 (1), p.47-58 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | One of the more significant indicators of neural age-related loss and disease is reduced temporal processing speed. It would, therefore, be useful to have an accurate and practical device that measures the full range of an individual's temporal processing abilities (characterized as the temporal contrast sensitivity function, TCSF). 70 subjects (15-84 yrs) were tested. A small tabletop device utilizing electronic control of light-emitting diodes (LEDs) was constructed that delivered a 1-degree, 660 nm test (the modulation depth of which could be adjusted directly by the subject) centered within a 10-degree 660 nm surround. The method provided a TCSF that had a shape consistent with past studies (peaking around 8 Hz). Also consistent with past work, the largest age-decline was found at the highest frequencies and for the central fovea (r = 0.47, p |
---|---|