Transcatheter closure with use of the SHSMA occluder in 180 patients with congenital heart defects: preliminary results

Transcatheter closure of congenital heart defects with the use of septal occluders has been widely accepted as a preferred treatment; however, the high cost of these devices limits their clinical application in some countries. Few clinical data are available regarding lower-cost products. Accordingl...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Texas Heart Institute journal 2010, Vol.37 (5), p.531-537
Hauptverfasser: Sun, Wan-Feng, Dong, Zhi-Feng, Gong, Kaizheng, Zhang, Guo-Pei, Cui, Ting, Xia, Yu-Dong, Dong, Jing, Shen, Yuan
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Transcatheter closure of congenital heart defects with the use of septal occluders has been widely accepted as a preferred treatment; however, the high cost of these devices limits their clinical application in some countries. Few clinical data are available regarding lower-cost products. Accordingly, we evaluated the efficacy and safety of the Chinese-made Shanghai Shape Memory Alloy (SHSMA) occluder in patients with congenital heart defects. From December 2001 through December 2008, a total of 180 patients with congenital heart defects (ages, 3-68 yr; mean age, 17.35 ± 13.22 yr) underwent transcatheter closure with use of the SHSMA occluder: 73 had atrial septal defects; 64, ventricular septal defects; 40, patent ductus arteriosus; and 3, complex congenital defects. The mean diameters of the defects were 20 ± 7.6 mm (atrial septal), 4.9 ± 2.1 mm (ventricular septal), and 5.6 ± 2.2 mm (patent ductus arteriosus). The procedural success rates were 98.6% for atrial defects, 98.4% for ventricular defects, and 100% for patent ductus arteriosus and for complex defects. The overall incidences of sequelae were 5.5%, 9.4%, 2.5%, and 0, respectively. Six months postprocedurally, complete occlusion was associated with a significant decrease in the right ventricular Tei index in atrial septal defect patients (P < 0.05) and with improvement of body mass index in 11 children. These results suggest that the SHSMA occluder is a safe, effective device for the transcatheter closure of congenital heart defects. For confirmation, a randomized controlled trial with more patients and a longer follow-up period is warranted.
ISSN:0730-2347