The effect of the trabecular microstructure on the pullout strength of suture anchors

Abstract This study investigates how the microstructural properties of trabecular bone affect suture anchor performance. Seven fresh-frozen humeri were tested for pullout strength with a 5 mm Arthrex Corkscrew in the greater tuberosity, lesser tuberosity, and humeral head. Micro-computed tomography...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of biomechanics 2010-07, Vol.43 (10), p.1953-1959
Hauptverfasser: Yakacki, Christopher M, Poukalova, Mariya, Guldberg, Robert E, Lin, Angela, Saing, Minn, Gillogly, Scott, Gall, Ken
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Abstract This study investigates how the microstructural properties of trabecular bone affect suture anchor performance. Seven fresh-frozen humeri were tested for pullout strength with a 5 mm Arthrex Corkscrew in the greater tuberosity, lesser tuberosity, and humeral head. Micro-computed tomography analysis was performed in the three regions of interest directly adjacent to individual pullout experiments. The morphometric properties of bone mineral density (BMD), structural model index (SMI), trabecular thickness (TbTh), trabecular spacing (TbS), trabecular number (TbN), and connectivity density were compared against suture anchor pullout strength. BMD ( r =0.64), SMI ( r =−0.81), and TbTh ( r =0.71) showed linear correlations to the pullout strength of the suture anchor with p -values
ISSN:0021-9290
1873-2380
DOI:10.1016/j.jbiomech.2010.03.013