Effect of lithotripter focal width on stone comminution in shock wave lithotripsy
Using a reflector insert, the original HM-3 lithotripter field at 20 kV was altered significantly with the peak positive pressure ( p + ) in the focal plane increased from 49 to 87 MPa while the −6 dB focal width decreased concomitantly from 11 to 4 mm. Using the original reflector, p + of 33 MPa wi...
Gespeichert in:
Veröffentlicht in: | The Journal of the Acoustical Society of America 2010-04, Vol.127 (4), p.2635-2645 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Using a reflector insert, the original HM-3 lithotripter field at 20 kV was altered significantly with the peak positive pressure
(
p
+
)
in the focal plane increased from 49 to 87 MPa while the −6 dB focal width decreased concomitantly from 11 to 4 mm. Using the original reflector,
p
+
of 33 MPa with a −6 dB focal width of 18 mm were measured in a pre-focal plane 15-mm proximal to the lithotripter focus. However, the acoustic pulse energy delivered to a 28-mm diameter area around the lithotripter axis was comparable
(
∼
120
mJ
)
. For all three exposure conditions, similar stone comminution
(
∼
70
%
)
was produced in a mesh holder of 15 mm after 250 shocks. In contrast, stone comminution produced by the modified reflector either in a 15-mm finger cot (45%) or in a 30-mm membrane holder (14%) was significantly reduced from the corresponding values (56% and 26%) produced by the original reflector (no statistically significant differences were observed between the focal and pre-focal planes). These observations suggest that a low-pressure/broad focal width lithotripter field will produce better stone comminution than its counterpart with high-pressure/narrow focal width under clinically relevant
in vitro
comminution conditions. |
---|---|
ISSN: | 0001-4966 1520-8524 |
DOI: | 10.1121/1.3308409 |