The potential use of Raman mapping to investigate in vitro deposition of combination pressurized metered-dose inhalers
Scanning near-infrared Raman microscopy has been used to map aerosol particulate deposits produced from pressurized metered-dose inhalers (pMDI). A commercially available combination asthma therapy pMDI (Ventide, Allen and Hanbury, UK), containing salbutamol and beclometasone dipropionate, was analy...
Gespeichert in:
Veröffentlicht in: | The AAPS journal 2004-10, Vol.6 (4), p.e32-44 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Scanning near-infrared Raman microscopy has been used to map aerosol particulate deposits produced from pressurized metered-dose inhalers (pMDI). A commercially available combination asthma therapy pMDI (Ventide, Allen and Hanbury, UK), containing salbutamol and beclometasone dipropionate, was analyzed by conventional in vitro quantitative analysis and scanning Raman microscopy. Raman maps, taken from Andersen cascade impactor plate stages 3 and 5 (over 100 x 100 microm areas) suggested good correlation with chemical analysis of the respective stages. Scanning Raman microscopy allows visual differentiation between formulation components (not possible using conventional imaging techniques), while potentially allowing chemical quantification. |
---|---|
ISSN: | 1550-7416 1550-7416 |
DOI: | 10.1208/aapsj060432 |