Pediatric organ dose measurements in axial and helical multislice CT
An anthropomorphic pediatric phantom (5-yr-old equivalent) was used to determine organ doses at specific surface and internal locations resulting from computed tomography (CT) scans. This phantom contains four different tissue-equivalent materials: Soft tissue, bone, brain, and lung. It was imaged o...
Gespeichert in:
Veröffentlicht in: | Medical physics (Lancaster) 2009-05, Vol.36 (5), p.1494-1499 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | An anthropomorphic pediatric phantom (5-yr-old equivalent) was used to determine organ doses at specific surface and internal locations resulting from computed tomography (CT) scans. This phantom contains four different tissue-equivalent materials: Soft tissue, bone, brain, and lung. It was imaged on a 64-channel CT scanner with three head protocols (one contiguous axial scan and two helical scans [
pitch
=
0.516
and 0.984]) and four chest protocols (one contiguous axial scan and three helical scans [
pitch
=
0.516
, 0.984, and 1.375]). Effective
mA
s
[
=
(
tube
current
×
rotation
time
)
/
pitch
]
was kept nearly constant at 200 effective
mA
s
for head and 290 effective
mA
s
for chest protocols. Dose measurements were acquired using thermoluminescent dosimeter powder in capsules placed at locations internal to the phantom and on the phantom surface. The organs of interest were the brain, both eyes, thyroid, sternum, both breasts, and both lungs. The organ dose measurements from helical scans were lower than for contiguous axial scans by 0% to 25% even after adjusting for equivalent effective
mA
s
. There was no significant difference
(
p
>
0.05
)
in organ dose values between the 0.516 and 0.984 pitch values for both head and chest scans. The chest organ dose measurements obtained at a pitch of 1.375 were significantly higher than the dose values obtained at the other helical pitches used for chest scans
(
p
<
0.05
)
. This difference was attributed to the automatic selection of the large focal spot due to a higher tube current value. These findings suggest that there may be a previously unsuspected radiation dose benefit associated with the use of helical scan mode during computed tomography scanning. |
---|---|
ISSN: | 0094-2405 2473-4209 0094-2405 |
DOI: | 10.1118/1.3101817 |