An ultrahigh resolution SPECT system for I-125 mouse brain imaging studies
This paper presents some initial experimental results obtained with a dual-head prototype single photon emission microscope system (SPEM) that is dedicated to mouse brain studies using I-125 labeled radiotracers. In particular, this system will be used for in vivo tacking of radiolabeled T cells in...
Gespeichert in:
Veröffentlicht in: | Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment Accelerators, spectrometers, detectors and associated equipment, 2009-03, Vol.600 (2), p.498-505 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | This paper presents some initial experimental results obtained with a dual-head prototype single photon emission microscope system (SPEM) that is dedicated to mouse brain studies using I-125 labeled radiotracers. In particular, this system will be used for
in vivo tacking of radiolabeled T cells in mouse brain. This system is based on the use of the intensified electron multiplying charge-coupled device (I-EMCCD) camera that offers the combination of an excellent intrinsic spatial resolution, a good signal-to-noise ratio, a large active area and a reasonable detection efficiency over an energy range between 27–140
keV. In this study, the dual-head SPEM system was evaluated using both resolution phantoms and a mouse with locally injected T cells labeled with I-125. It was demonstrated that for a relatively concentrated source object, the current dual-head SPEM system is capable of visualizing the tiny amount of radioactivity (∼12
nCi) carried by a very small number ( |
---|---|
ISSN: | 0168-9002 1872-9576 |
DOI: | 10.1016/j.nima.2008.11.149 |