Falsely raised TSH levels due to human anti-mouse antibody interfering with thyrotropin assay

The case of a 39-year-old woman who was referred for weight gain and amenorrhoea is reported. Laboratory evaluation showed high levels of thyroid-stimulating hormone (TSH). The patient was started on increasing doses of levothyroxine for subclinical hypothyroidism. TSH remained persistently raised a...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Postgraduate medical journal 2006-11, Vol.82 (973), p.e27-e27
Hauptverfasser: Krishnan, S G S, Pathalapati, R, Kaplan, L, Cobbs, R K
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The case of a 39-year-old woman who was referred for weight gain and amenorrhoea is reported. Laboratory evaluation showed high levels of thyroid-stimulating hormone (TSH). The patient was started on increasing doses of levothyroxine for subclinical hypothyroidism. TSH remained persistently raised and the patient became thyrotoxic. Evaluation at another laboratory showed normal levels of TSH, raising the possibility of interfering substances. TSH levels were normalised with the addition of mouse serum to the patient’s sample, confirming the presence of human anti-mouse antibodies as the interfering substance in the TSH assay.
ISSN:0032-5473
1469-0756
DOI:10.1136/pmj.2006.049809