Multicentre evaluation of the Boehringer Mannheim/Hitachi 917 analysis system
The new selective access analysis system BM/Hitachi 917 was evaluated in an international multicentre study, mainly according to the ECCLS protocol for the evaluation of analysers in clinical chemistry. Forty-three different analytes, covering 56 different methods--enzymes, substrates, electrolytes,...
Gespeichert in:
Veröffentlicht in: | Journal of analytical methods in chemistry 2000, Vol.22 (3), p.65-81 |
---|---|
Hauptverfasser: | , , , , , , , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The new selective access analysis system BM/Hitachi 917 was evaluated in an international multicentre study, mainly according to the ECCLS protocol for the evaluation of analysers in clinical chemistry. Forty-three different analytes, covering 56 different methods--enzymes, substrates, electrolytes, specific proteins, drugs and urine applications--were tested in seven European clinical chemistry laboratories. Additionally, the practicability of the BM/ Hitachi 917 was tested according to a standardized questionnaire. Within-run CVs (median of 3 days) for enzymes, substrates and electrolytes were |
---|---|