Comparison of estimated glomerular filtration rate with routine creatinine clearance using a kinetic alkaline picrate assay from Olympus Diagnostica

The Passing Bablok linear regression equation 11 with creatinine clearance as x and eGFR as y was y=0.613x+2.9 with the 95% CIs for the slope and intercept being 0.55 to 0.682 and -0.7 to +6.2, respectively. [...]the slope is significantly different from 1.0 and is consistent with a bias of approxim...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of clinical pathology 2007-06, Vol.60 (6), p.732-733
Hauptverfasser: Twomey, Patrick J, Pledger, Derek R
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The Passing Bablok linear regression equation 11 with creatinine clearance as x and eGFR as y was y=0.613x+2.9 with the 95% CIs for the slope and intercept being 0.55 to 0.682 and -0.7 to +6.2, respectively. [...]the slope is significantly different from 1.0 and is consistent with a bias of approximately 38.7% between the two methods. [...]the routine use of eGFR would result in fewer individual patients having a CI that crosses one of the CKD classification cut-offs compared with the CCT.
ISSN:0021-9746
1472-4146
DOI:10.1136/jcp.2005.033472