Comparison of estimated glomerular filtration rate with routine creatinine clearance using a kinetic alkaline picrate assay from Olympus Diagnostica
The Passing Bablok linear regression equation 11 with creatinine clearance as x and eGFR as y was y=0.613x+2.9 with the 95% CIs for the slope and intercept being 0.55 to 0.682 and -0.7 to +6.2, respectively. [...]the slope is significantly different from 1.0 and is consistent with a bias of approxim...
Gespeichert in:
Veröffentlicht in: | Journal of clinical pathology 2007-06, Vol.60 (6), p.732-733 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The Passing Bablok linear regression equation 11 with creatinine clearance as x and eGFR as y was y=0.613x+2.9 with the 95% CIs for the slope and intercept being 0.55 to 0.682 and -0.7 to +6.2, respectively. [...]the slope is significantly different from 1.0 and is consistent with a bias of approximately 38.7% between the two methods. [...]the routine use of eGFR would result in fewer individual patients having a CI that crosses one of the CKD classification cut-offs compared with the CCT. |
---|---|