Clinical results with the Trabectome, a novel surgical device for treatment of open-angle glaucoma
To describe treatment outcomes after Trabectome surgery in an initial series of 101 patients with open-angle glaucoma. A 19-gauge microelectrosurgical device enabled ab interno removal of a strip of trabecular meshwork and inner wall of Schlemm's canal under gonioscopic control with continual i...
Gespeichert in:
Veröffentlicht in: | Transactions of the American Ophthalmological Society 2006, Vol.104, p.40-50 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | To describe treatment outcomes after Trabectome surgery in an initial series of 101 patients with open-angle glaucoma.
A 19-gauge microelectrosurgical device enabled ab interno removal of a strip of trabecular meshwork and inner wall of Schlemm's canal under gonioscopic control with continual infusion and foot-pedal control of aspiration and electrosurgery. A smooth, pointed ceramic-coated insulating footplate was inserted into Schlemm's canal to act as a guide within the canal and to protect adjacent structures from mechanical or heat injury during ablation of a 30- to 90-degree arc of angle tissue.
Mean preoperative intraocular pressure (IOP) in the initial 101 patients was 27.6 +/- 7.2 mm Hg. Thirty months postoperatively, mean IOP was 16.3 +/- 3.3 mm Hg (n = 11). The mean percentage drop over the whole course of follow-up was 40%. At all times postoperatively, the absolute and percent decrease in IOP from preoperative levels were statistically significant (paired t test, P < .0001). Overall success (IOP |
---|---|