Endocardial and epicardial steroid lead pacing in the neonatal and paediatric age group
Aim: To compare the performance of steroid eluting epicardial and endocardial leads in infants and children requiring permanent pacing. Methods: Evaluation of pacing and sensing characteristics, impedances, and longevity of 159 steroid eluting leads implanted in 95 children. Group A consisted of 24...
Gespeichert in:
Veröffentlicht in: | British heart journal 2002-10, Vol.88 (4), p.392-396 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Aim: To compare the performance of steroid eluting epicardial and endocardial leads in infants and children requiring permanent pacing. Methods: Evaluation of pacing and sensing characteristics, impedances, and longevity of 159 steroid eluting leads implanted in 95 children. Group A consisted of 24 children weighing less than 15 kg with 15 endocardial leads (five atrial, 10 ventricular) and 19 epicardial leads (five atrial, 14 ventricular). Group B consisted of 71 children weighing more than 15 kg with 106 endocardial leads (56 atrial, 58 ventricular) and 19 epicardial leads (nine atrial, 10 ventricular). Results: Group A: Stimulation thresholds were lower for ventricular endocardial leads at implant (mean (SD) 0.84 (0.54) v 1.59 (0.64) V, p < 0.014) and at two year follow up (ventricular 0.64 (0.24) v 1.65 (0.69) V, p < 0.003). Impedance and sensing thresholds did not differ significantly at implant and follow up. Group B: Stimulation thresholds were lower for ventricular endocardial leads at implant (0.72 (0.48) v 1.48 (0.58) V, p < 0.001) and at follow up (0.88 (0.46) v 1.55 (0.96) V, p < 0.009). Impedance did not differ. Sensing thresholds were also better for ventricular endocardial leads at follow up (9.1 (5.2) v 14.2 (6.4) mV, p < 0.02). Complications requiring intervention occurred in both groups (n = 7 for endocardial v n = 18 for epicardial leads). Conclusions: Endocardial and epicardial steroid eluting leads have comparable performance in the paediatric population. |
---|---|
ISSN: | 1355-6037 0007-0769 1468-201X |
DOI: | 10.1136/heart.88.4.392 |