Performance of an interstitial glucose monitoring device in patients with type 1 diabetes during haemodialysis
ABSTRACT Background The use of interstitial glucose monitoring devices such as flash glucose monitoring has been shown to be beneficial in patients with type 1 diabetes mellitus (T1DM). However, these devices have been little studied in patients with diabetes treated by chronic haemodialysis (HD). M...
Gespeichert in:
Veröffentlicht in: | Clinical kidney journal 2024-09, Vol.17 (9), p.sfae045 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | ABSTRACT
Background
The use of interstitial glucose monitoring devices such as flash glucose monitoring has been shown to be beneficial in patients with type 1 diabetes mellitus (T1DM). However, these devices have been little studied in patients with diabetes treated by chronic haemodialysis (HD).
Methods
The goal of this prospective, observational, multicentric study was to evaluate the analytical performance of the FreeStyle Libre 2 (FSL2) sensor in T1DM patients during HD sessions. During three HD sessions, interstitial fluid glucose (ISFG) concentrations given by the FSL2 were compared every 15 minutes with blood glucose (BG) concentrations obtained simultaneously. BG concentrations were measured by two different glucometers: the Accu-Chek Guide and StatStrip meters.
Results
Twelve HD patients were included, with a mean age of 54 ± 11 years and a mean diabetes duration of 36.5 ± 11.6 years. Dialysis vintage was 35 ± 22 months. A total of 565 pairs of ISFG/BG values were available for analysis. The mean absolute relative difference, defined as the mean of the absolute relative differences between the ISFG and BG measurements, was 17.4% and 20.9% when the ISFG was compared with the StatStrip meter or Accu-Chek Guide, respectively. Interstitial results tend to underestimate blood results, but all values were classified as having clinically acceptable error. The differences observed remained stable during the dialysis session and were not associated with the ultrafiltration rate.
Conclusion
Use of the FSL2 interstitial glucose monitoring device in HD patients with T1DM is clinically acceptable, even though the accuracy of the device is generally poorer than in studies including non-dialysis patients.