Atrioventricular Synchrony Delivered by a Dual-Chamber Leadless Pacemaker System
A dual-chamber leadless pacemaker system has been designed for atrioventricular synchronous pacing using wireless, beat-to-beat, implant-to-implant (i2i) communication between distinct atrial and ventricular leadless pacemakers. The atrioventricular synchrony achieved across various ambulatory scena...
Gespeichert in:
Veröffentlicht in: | Circulation (New York, N.Y.) N.Y.), 2024-08, Vol.150 (6), p.439-450 |
---|---|
Hauptverfasser: | , , , , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A dual-chamber leadless pacemaker system has been designed for atrioventricular synchronous pacing using wireless, beat-to-beat, implant-to-implant (i2i) communication between distinct atrial and ventricular leadless pacemakers. The atrioventricular synchrony achieved across various ambulatory scenarios has yet to be systematically evaluated.
A prospective, single-arm, unblinded, multicenter, international clinical trial of the leadless pacemaker system was conducted in patients with a conventional dual-chamber pacing indication enrolled from February 2022 to March 2023. Leadless pacemaker systems were implanted, and 12-lead Holter electrocardiographic recordings were collected 3 months after implantation over various postures/activities: sitting, supine, left lateral recumbent, right lateral recumbent, standing, normal walk, and fast walk. An independent Holter core laboratory performed a manual adjudication of the percent of atrioventricular synchronous beats using the standard 300-millisecond PR interval limit. Atrium-to-ventricle and ventricle-to-atrium i2i communication success rates were also assessed. Post hoc summary statistics describing the relationships between atrioventricular synchrony and i2i success, posture/activity, implantation indication, atrioventricular event, and heart rate were calculated.
In the evaluable population (n=384 of 464 enrolled [83%]; 61% male; age, 70 years; weight, 82 kg; 60% ejection fraction; 95% of beats evaluable), the mean atrioventricular synchrony of 98% of beats observed across all postures using the standard 300-millisecond limit was greater than both atrial-to-ventricular i2i (94%) and ventricular-to-atrial i2i (94%;
95% of evaluable beats across all postures/activities, implantation indications, atrioventricular paced/sensed event combinations, and heart rate ranges (including >100 bpm).
This dual-chamber leadless pacemaker system demonstrated atrioventricular synchrony in 98% of evaluable beats at 3 months after implantation. Atrioventricular synchrony was maintained across postures/activities and remained robust for heart rates >100 bpm. |
---|---|