Comparative Profiles of the WATCHMAN™ and Amplatzer™ Cardiac Plug/Amplatzer™ Amulet™ Devices for Left Atrial Appendage Closure in Non-valvular Atrial Fibrillation: A Comprehensive Systematic Review and Meta-analysis
Atrial fibrillation (AF) is a prevalent cardiac arrhythmia marked by irregular and frequent tachycardic rhythms in the atria, affecting 1%-2% of the general population. The WATCHMAN™ device from Boston Scientific (Marlborough, MA, USA) and the Amplatzer™ Amulet™ device from Abbott (Chicago, IL, USA)...
Gespeichert in:
Veröffentlicht in: | The Journal of innovations in cardiac rhythm management (Print) 2024-06, Vol.15 (6), p.5917-5929 |
---|---|
Hauptverfasser: | , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Atrial fibrillation (AF) is a prevalent cardiac arrhythmia marked by irregular and frequent tachycardic rhythms in the atria, affecting 1%-2% of the general population. The WATCHMAN™ device from Boston Scientific (Marlborough, MA, USA) and the Amplatzer™ Amulet™ device from Abbott (Chicago, IL, USA) are two devices used globally for left atrial appendage closure (LAAC) in non-valvular AF. A systematic search was conducted in PubMed, the Cochrane Library, and Elsevier's ScienceDirect literature databases to identify studies comparing the WATCHMAN™ procedure with Amulet™ device implantation for LAAC in patients with AF. The analyses were conducted using the random-effects model. A total of 20 studies were identified, with 18 falling into the category of observational studies and 2 being randomized controlled trials. A total of 6310 participants were included in this meta-analysis, with 3198 individuals (50.68%) assigned to the WATCHMAN™ procedure group and 3112 individuals (49.32%) allocated to the Amplatzer™ Cardiac Plug (ACP) group. The analysis revealed a higher risk of stroke associated with the WATCHMAN™ technique (relative risk [RR], 1.14), albeit without statistical significance. Conversely, the WATCHMAN™ approach led to a significantly lower risk of cardiac death (RR, 0.44;
= .04). Notably, the risks of all-cause mortality (RR, 0.89; 95% confidence interval [CI], 0.73-1.08;
= 0%;
= .25) and major bleeding (RR, 0.93; 95% CI, 0.65-1.33;
= 31%;
= .70) were clinically reduced with the WATCHMAN™ procedure, although statistical significance was not achieved. Compared to Amulet™ device implantation, WATCHMAN™ device implantation decreased the risk of cardiac mortality, while the risks of stroke, systemic embolism, all-cause mortality, and major bleeding were not statistically significant. |
---|---|
ISSN: | 2156-3977 2156-3993 |
DOI: | 10.19102/icrm.2024.15061 |