Quartz crystal microbalance–based aptasensor integrated with magnetic pre-concentration system for detection of Listeria monocytogenes in food samples
A fast and accurate identification of Listeria monocytogenes. A new quartz crystal microbalance (QCM) aptasensor was designed for the specific and rapid detection of L. monocytogenes . Before detection of the target bacterium from samples in the QCM aptasensor, a magnetic pre-enrichment system was u...
Gespeichert in:
Veröffentlicht in: | Mikrochimica acta (1966) 2024-05, Vol.191 (5), p.235-235, Article 235 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A fast and accurate identification of
Listeria monocytogenes.
A new quartz crystal microbalance (QCM) aptasensor was designed for the specific and rapid detection of
L. monocytogenes
. Before detection of the target bacterium from samples in the QCM aptasensor, a magnetic pre-enrichment system was used to eliminate any contaminant in the samples. The prepared magnetic system was characterized using ATR-FTIR, SEM, VSM, BET, and analytical methods. The saturation magnetization values of the Fe
3
O
4
, Fe
3
O
4
@PDA, and Fe
3
O
4
@PDA@DAPEG particles were 57.2, 40.8, and 36.4 emu/g, respectively. The same aptamer was also immobilized on the QCM crystal integrated into QCM flow cell and utilized to quantitatively detect
L. monocytogenes
cells from the samples. It was found that a specific aptamer-magnetic pre-concentration system efficiently captured
L. monocytogenes
cells in a short time (approximately 10 min). The Fe
3
O
4
@PDA@DA-PEG-Apt particles provided selective isolation of
L. monocytogenes
from the bacteria-spiked media up to 91.8%. The immobilized aptamer content of the magnetic particles was 5834 µg/g using 500 ng Apt/mL. The QCM aptasensor showed a very high range of analytical performance to the target bacterium from 1.0 × 10
2
and 1.0 × 10
7
CFU/mL. The limit of detection (LOD) and limit of quantitation (LOQ) were 148 and 448 CFU/mL, respectively, from the feeding of the QCM aptasensor flow cell with the eluent of the magnetic pre-concentration system. The reproducibility of the aptasensor was more than 95%. The aptasensor was very specific to
L. monocytogenes
compared to the other
Listeria species
(i.e.,
L
.
ivanovii
,
L
.
innocua
, and
L. seeligeri
) or other tested bacteria such as
Staphylococcus aureus
,
Escherichia coli
, and
Bacillus subtilis
. The QCM aptasensor was regenerated with NaOH solution, and the system was reused many times.
Graphical Abstract |
---|---|
ISSN: | 0026-3672 1436-5073 |
DOI: | 10.1007/s00604-024-06307-2 |