Multi-window approach enables two-fold improvement in OCT axial resolution with strong side-lobe suppression and improved phase sensitivity
A common processing approach for optical coherence tomography (OCT) uses a window function (e.g., Hann or rectangular window) for spectral shaping prior to calculating the Fourier transform. Here we build on a multi-window approach [Opt. Express8, 5267 (2017)10.1364/BOE.8.005267] that enables improv...
Gespeichert in:
Veröffentlicht in: | Biomedical optics express 2023-12, Vol.14 (12), p.6301-6316 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A common processing approach for optical coherence tomography (OCT) uses a window function (e.g., Hann or rectangular window) for spectral shaping prior to calculating the Fourier transform. Here we build on a multi-window approach [Opt. Express8, 5267 (2017)10.1364/BOE.8.005267] that enables improved resolution while still suppressing side-lobe intensity. The shape of the window function defines the trade-off between main-lobe width (resolution) and side-lobe intensity. We have extended the approach to include the interferometric phase for phase-sensitive applications like vibrometry and Doppler OCT. Using the Hann window as a reference, we show that 11 Taylor windows are sufficient to achieve 50% improvement in axial resolution, -31 dB side-lobe intensity, and 20% improvement in phase sensitivity with low computational cost. |
---|---|
ISSN: | 2156-7085 2156-7085 |