Initial experience with an electron FLASH research extension (FLEX) for the Clinac system

Purpose Radiotherapy delivered at ultra‐high‐dose‐rates (≥40 Gy/s), that is, FLASH, has the potential to effectively widen the therapeutic window and considerably improve the care of cancer patients. The underlying mechanism of the FLASH effect is not well understood, and commercial systems capable...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of Applied Clinical Medical Physics 2024-02, Vol.25 (2), p.e14159-n/a
Hauptverfasser: Oh, Kyuhak, Gallagher, Kyle J., Hyun, Megan, Schott, Diane, Wisnoskie, Sarah, Lei, Yu, Hendley, Samuel, Wong, Jeffrey, Wang, Shuo, Graff, Brendan, Jenkins, Christopher, Rutar, Frank, Ahmed, Md, McNeur, Joshua, Taylor, Jeffrey, Schmidt, Marty, Senadheera, Lasitha, Smith, Wendy, Umstadter, Donald, Lele, Subodh M., Dai, Ran, Jianghu (James), Dong, Yan, Ying, Su‐min, Zhou
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Purpose Radiotherapy delivered at ultra‐high‐dose‐rates (≥40 Gy/s), that is, FLASH, has the potential to effectively widen the therapeutic window and considerably improve the care of cancer patients. The underlying mechanism of the FLASH effect is not well understood, and commercial systems capable of delivering such dose rates are scarce. The purpose of this study was to perform the initial acceptance and commissioning tests of an electron FLASH research product for preclinical studies. Methods A linear accelerator (Clinac 23EX) was modified to include a non‐clinical FLASH research extension (the Clinac‐FLEX system) by Varian, a Siemens Healthineers company (Palo Alto, CA) capable of delivering a 16 MeV electron beam with FLASH and conventional dose rates. The acceptance, commissioning, and dosimetric characterization of the FLEX system was performed using radiochromic film, optically stimulated luminescent dosimeters, and a plane‐parallel ionization chamber. A radiation survey was conducted for which the shielding of the pre‐existing vault was deemed sufficient. Results The Clinac‐FLEX system is capable of delivering a 16 MeV electron FLASH beam of approximately 1 Gy/pulse at isocenter and reached a maximum dose rate >3.8 Gy/pulse near the upper accessory mount on the linac gantry. The percent depth dose curves of the 16 MeV FLASH and conventional modes for the 10 × 10 cm2 applicator agreed within 0.5 mm at a range of 50% of the maximum dose. Their respective profiles agreed well in terms of flatness but deviated for field sizes >10 × 10 cm2. The output stability of the FLASH system exhibited a dose deviation of
ISSN:1526-9914
1526-9914
DOI:10.1002/acm2.14159