Objective autorefraction in posterior chamber pseudophakia
Automated refraction with the Canon RK-1 Autoref keratometer was evaluated in 110 eyes (110 patients) six to eight weeks after they had undergone extracapsular cataract extraction with posterior chamber intraocular lens implantation and achieved a best corrected visual acuity of at least 6/12. Autor...
Gespeichert in:
Veröffentlicht in: | British journal of ophthalmology 1990-12, Vol.74 (12), p.731-733 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Automated refraction with the Canon RK-1 Autoref keratometer was evaluated in 110 eyes (110 patients) six to eight weeks after they had undergone extracapsular cataract extraction with posterior chamber intraocular lens implantation and achieved a best corrected visual acuity of at least 6/12. Autorefraction readings were obtained in 100 (91%) of these eyes. The agreement between autorefraction and clinical refraction data was 98% for spherical equivalence less than 0.51 dioptres (D), 95% for sphere power less than 0.51D, 94% for cylinder power less than 0.51D, and 85% for cylinder axis less than 11 degrees. Autorefraction can provide acceptably accurate postoperative refraction values in pseudophakic eyes. |
---|---|
ISSN: | 0007-1161 1468-2079 |
DOI: | 10.1136/bjo.74.12.731 |