Four-color fluorescence cross-correlation spectroscopy with one laser and one camera

The diffusion and reorganization of phospholipids and membrane-associated proteins are fundamental for cellular function. Fluorescence cross-correlation spectroscopy (FCCS) measures diffusion and molecular interactions at nanomolar concentration in biological systems. We have developed an economical...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Biomedical optics express 2023-07, Vol.14 (7), p.3812-3827
Hauptverfasser: Gandhi, Sonali A, Sanders, Matthew A, Granneman, James G, Kelly, Christopher V
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The diffusion and reorganization of phospholipids and membrane-associated proteins are fundamental for cellular function. Fluorescence cross-correlation spectroscopy (FCCS) measures diffusion and molecular interactions at nanomolar concentration in biological systems. We have developed an economical method to simultaneously monitor diffusion and complexation with the use of super-continuum laser and spectral deconvolution from a single detector. Customizable excitation wavelengths were chosen from the wide-band source and spectral fitting of the emitted light revealed the interactions for up to four chromatically overlapping fluorophores simultaneously. This method was applied to perform four-color FCCS that we demonstrated with polystyrene nanoparticles, lipid vesicles, and membrane-bound molecules. Up to four individually customizable excitation channels were selected from the broad-spectrum fiber laser to excite the diffusers within a diffraction-limited spot. The fluorescence emission passed through a cleanup filter and a dispersive prism prior to being collected by a sCMOS or EMCCD camera with up to 1.8 kHz frame rates. The emission intensity versus time of each fluorophore was extracted through a linear least-square fitting of each camera frame and temporally correlated via custom software. Auto- and cross-correlation functions enabled the measurement of the diffusion rates and binding partners. We have measured the induced aggregation of nanobeads and lipid vesicles in solution upon increasing the buffer salinity. Because of the adaptability of investigating four fluorophores simultaneously with a cost-effective method, this technique will have wide application for examining macromolecular complex formation in model and living systems.
ISSN:2156-7085
2156-7085
DOI:10.1364/BOE.486937