Untangling a Knotted Angiographic Catheter Using a Balloon Catheter: A Case Report
When a 66-year-old man with hepatocellular carcinoma underwent an angiographic examination, a 4-Fr catheter was inserted from the right femoral artery. It became tightly knotted in the descending aorta. To untangle the knotted catheter, a noncompliant balloon catheter was delivered into the knotted...
Gespeichert in:
Veröffentlicht in: | Interventional Radiology 2023/07/01, Vol.8(2), pp.80-82 |
---|---|
Hauptverfasser: | , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | When a 66-year-old man with hepatocellular carcinoma underwent an angiographic examination, a 4-Fr catheter was inserted from the right femoral artery. It became tightly knotted in the descending aorta. To untangle the knotted catheter, a noncompliant balloon catheter was delivered into the knotted loop from the contralateral femoral artery. After the balloon catheter was inflated from the inside of the knotted loop, the knot became loose. Finally, the knotted catheter was untangled. Subsequently, the remainder of the examination was performed as initially planned. |
---|---|