Evaluation of a Fourth-Generation Subcutaneous Real-Time Continuous Glucose Monitor (CGM) in Individuals With Diabetes on Peritoneal Dialysis
To evaluate the performance of a real-time continuous glucose monitor (CGM) in individuals with diabetes on peritoneal dialysis (PD). Thirty participants with type 2 diabetes on continuous ambulatory PD wore a Guardian Sensor 3 on the upper arm paired with Guardian Connect for 14 days. We compared C...
Gespeichert in:
Veröffentlicht in: | Diabetes care 2023-06, Vol.46 (6), p.1191-1195 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | To evaluate the performance of a real-time continuous glucose monitor (CGM) in individuals with diabetes on peritoneal dialysis (PD).
Thirty participants with type 2 diabetes on continuous ambulatory PD wore a Guardian Sensor 3 on the upper arm paired with Guardian Connect for 14 days. We compared CGM readings against Yellow Springs Instrument (YSI) venous glucose during an 8-h in-clinic session with glucose challenge.
The mean absolute relative difference (MARD) was 10.4% (95% CI 9.6, 11.7) from 941 CGM-YSI matched pairs; 81.3% of readings were within %15/15 of YSI values in the full glycemic range. Consensus error grid analysis showed 99.9% of sensor values in zones A and B. There were no correlations between pH, uremia, hydration status, and MARD.
We showed satisfactory performance of a real-time CGM sensor in PD patients with diabetes, supporting future use to facilitate treatment decisions. |
---|---|