Reliability and Validity of Shoulder and Handgrip Strength Testing

This study aimed to estimate the intra- and inter-rater reliability of the JTECH computerized, wireless apparatus and its validity (compared to established devices) for measurements of maximal shoulder isometric strength and handgrip strength among healthy adults with no shoulder pathology. Twenty h...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Physiotherapy Canada 2023-02, Vol.75 (1), p.65-71
Hauptverfasser: Biasini, Nicholas Robert, Bannon, Brian, Pellegrino, Michael, Qaderi, Abdullah, Trinh, William, Switzer-McIntyre, Sharon, Reid, W Darlene, Kasawara, Karina Tamy
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:This study aimed to estimate the intra- and inter-rater reliability of the JTECH computerized, wireless apparatus and its validity (compared to established devices) for measurements of maximal shoulder isometric strength and handgrip strength among healthy adults with no shoulder pathology. Twenty healthy young adults had shoulder strength tested with JTECH and Micro-FET2 hand-held dynamometers, and the handgrip strength was tested using JTECH and Jamar handgrip dynamometers. Assessments were performed by the same rater to determine intra-rater reliability and convergent validity, at least two days apart, while on a third visit, the other rater performed measures to determine inter-rater reliability. The JTECH computerized, wireless devices demonstrated good to excellent intra-rater reliability (ICCs (2,1) = 0.78-0.97) and inter-rater reliability for strength measures (ICCs (2,1) = 0.76-0.95). The JTECH computerized device compared to the Micro-FET2 hand-held dynamometer showed substantial concurrent validity for shoulder flexion (  = 0.87), extension (  = 0.87), abduction (  = 0.88), and adduction (  = 0.85). Substantial concurrent validity was shown for the JTECH computerized device and Jamar handgrip dynamometers (  = 0.92). The JTECH computerized, wireless devices demonstrated high intra- and inter-rater reliability and substantial concurrent validity for shoulder isometric strength and handgrip strength in healthy adults.
ISSN:0300-0508
1708-8313