The NIST Vacuum Double-Crystal Spectrometer: A Tool for SI-Traceable Measurement of X-Ray Emission Spectra
The NIST Vacuum Double-Crystal Spectrometer (VDCS) has been modernized and is now capable of recording reference-free wavelength-dispersive spectra in the 2 keV to 12 keV x-ray energy range. The VDCS employs crystals in which the lattice spacings are traceable to the definition of the meter through...
Gespeichert in:
Veröffentlicht in: | Journal of research of the National Institute of Standards and Technology 2021, Vol.126, p.126049-29, Article 126049 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The NIST Vacuum Double-Crystal Spectrometer (VDCS) has been modernized and is now capable of recording reference-free wavelength-dispersive spectra in the 2 keV to 12 keV x-ray energy range. The VDCS employs crystals in which the lattice spacings are traceable to the definition of the meter through x-ray optical interferometry with a relative uncertainty ﹤10-⁸. VDCS wavelength determination relies upon precision angle difference measurements for which the encoders of the rotation stages have been calibrated using the circle closure method for accurate, absolute angle measurement. The new vacuum-compatible area detector allows quantification of the aberration functions contributing to the observed line shape and in situ alignment of the crystal optics. This latter procedure is augmented with the use of a thin lamella as the first crystal. With these new techniques, x-ray spectra are registered with the VDCS on an absolute energy scale with a relative uncertainty of 10-⁶. |
---|---|
ISSN: | 1044-677X 2165-7254 2165-7254 |
DOI: | 10.6028/jres.126.049 |