Differences in Electrical Delay Between the Mid-septum and Apex with the Right Ventricular Lead: Novel Implications for Pacemaker Sensing
Ventricular sensing relies on the analysis of a local intracardiac electrogram in reference to the QRS on the surface electrocardiogram. If both signals do not coincide in time, there is a delay in sensing intrinsic ventricular activity. We evaluated possible differences in the electrical delay betw...
Gespeichert in:
Veröffentlicht in: | The Journal of innovations in cardiac rhythm management (Print) 2023-03, Vol.14 (3), p.5372-5378 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Ventricular sensing relies on the analysis of a local intracardiac electrogram in reference to the QRS on the surface electrocardiogram. If both signals do not coincide in time, there is a delay in sensing intrinsic ventricular activity. We evaluated possible differences in the electrical delay between the mid-septum and apex as determined by the right ventricular (RV) lead position using a pacing system analyzer (PSA) during conventional pacemaker implantation. Patients without significant heart disease and intrinsic atrioventricular conduction underwent their first Medtronic (Minneapolis, MN, USA) or Abbott (Chicago, IL, USA) dual-chamber pacemaker implantation with the RV lead first positioned at the apex and then subsequently at the mid-septum. Real-time ventricular sensing data were obtained through PSA to determine the electrical delay Q-VS value as the time difference between the QRS and the released RV-sensed event marker "VS." Among 212 patients, 139 had narrow QRS and 73 had complete right bundle branch block (RBBB). Overall, both narrow QRS and RBBB patients exhibited shorter Q-VS lengths at the mid-septum compared to the apex (50.4 ± 24.2 ms and 66.7 ± 32.3 ms vs. 63.9 ± 27.6 ms and 71.7 ± 32.2 ms;
< .0001 and
< .001, respectively). The Q-VS in patients with Abbott devices was significantly shorter compared to that in patients with Medtronic devices at both the mid-septum and the apex in both patient groups (
< .0001). In conclusion, RV lead positioning at the mid-septum is associated with a shorter electrical delay compared to positioning at the apex in both narrow QRS and RBBB patients. |
---|---|