Standing-wave and RF penetration artifacts caused by elliptic geometry: an electrodynamic analysis of MRI

Motivated by the observation that the diagonal pattern of intensity nonuniformity usually associated with linearly polarized radio-frequency (RF) coils is often present in neurological scans using circularly polarized coils, a theoretical analysis has been conducted of the intensity nonuniformity in...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on medical imaging 1998-08, Vol.17 (4), p.653-662
Hauptverfasser: Sled, J.G., Pike, G.B.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Motivated by the observation that the diagonal pattern of intensity nonuniformity usually associated with linearly polarized radio-frequency (RF) coils is often present in neurological scans using circularly polarized coils, a theoretical analysis has been conducted of the intensity nonuniformity inherent in imaging an elliptically shaped object using 1.5-T magnets and circularly polarized RF coils. This first principle analysis clarifies, for the general case of conducting objects, the relationship between the excitation field and the reception sensitivity of circularly and linearly polarized coils. The results, validated experimentally using a standard spin-echo imaging sequence and an in vivo B/sub 1/ field mapping technique, are shown to be accurate to within 1%-2% root mean square, suggesting that these electromagnetic interactions with the object account for most of the intensity nonuniformity observed.
ISSN:0278-0062
1558-254X
DOI:10.1109/42.730409