Absorbed doses with intraoral radiography : function of various technical parameters

Both the maxillary incisor and mandibular molar regions of a Rando phantom were radiographed using specific technical parameters. The parotid gland and thyroid gland doses were measured using thermoluminescence dosimetry. Tube voltage, total filtration, mAs, collimation, and cone length were varied...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Oral surgery, oral medicine, oral pathology oral medicine, oral pathology, 1993-10, Vol.76 (4), p.519-524
Hauptverfasser: HAYAKAWA, Y, FUJIMORI, H, KUROYANAGI, K
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Both the maxillary incisor and mandibular molar regions of a Rando phantom were radiographed using specific technical parameters. The parotid gland and thyroid gland doses were measured using thermoluminescence dosimetry. Tube voltage, total filtration, mAs, collimation, and cone length were varied while keeping the other factors constant. Increased tube voltage or total filtration generally resulted in a slightly increased absorbed dose. Absorbed dose was decreased as the beam was more tightly collimated or the cone length was increased.
ISSN:0030-4220
1878-2175