In Vitro Characterization of the RP-6 Dialyzer: Co-Current and Counter-Current Clearance as a Function of Ultrafiltration
Analytical and graphical techniques are described for presenting clearance data as a function of ultrafiltration; thus for the first time comparison of clearance values between dialyzers can be made unambiguously. This simple method uses a hand calculator to determine three coefficients. Clearance h...
Gespeichert in:
Veröffentlicht in: | Journal of dialysis 1979, Vol.3 (2-3), p.237-250 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Analytical and graphical techniques are described for presenting clearance data as a function of ultrafiltration; thus for the first time comparison of clearance values between dialyzers can be made unambiguously. This simple method uses a hand calculator to determine three coefficients. Clearance has been shown to be accurately described by the equation C = α1 + α2Qv + α3 (1/QBi). This formula permits construction of easily read families of curves which can predict clearance values under most clinical situations. The RP-6 dialyzer is characterized by these methods for clearance of urea, glucose and vitamin B12 for counter-current and co-current modes. |
---|---|
ISSN: | 0886-022X 0362-8558 1525-6049 |