Effect of aggregates on albumin standardization

In order to investigate the consequences of presence of aggregates in human albumin standards, pools of monomer, dimer and polymer albumin were prepared and quantitated by three total-protein methods (biuret, Folin-Lowry and spectrophotometry at 279 nm) and by four different albumin methods (dye-bin...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Scandinavian journal of clinical and laboratory investigation 1979-01, Vol.39 (8), p.751-757
Hauptverfasser: Blaabjerg, O., Petersen, P. Hyltoft
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:In order to investigate the consequences of presence of aggregates in human albumin standards, pools of monomer, dimer and polymer albumin were prepared and quantitated by three total-protein methods (biuret, Folin-Lowry and spectrophotometry at 279 nm) and by four different albumin methods (dye-binding by bromcresol green, electroimmunoassay, radial immunodiffusion and automated immunoprecipitation). Biuret was chosen as the reference method and the monomer was used as the standard in all methods. Both the total-protein and albumin methods gave values for aggregated albumin different from the biuret values. The maximal bias occurred in radial immunodiffusion where quantitation of the dimer and polymer pools gave only 67% and 48% of the biuret values, respectively. In five commercial albumin preparations investigated, the content of di- and polymer varied from 3 to 34%. Uncritical use of albumin preparations in standardization may thus introduce bias in the measurements.
ISSN:0036-5513
1502-7686
DOI:10.3109/00365517909108167