Establishing Updated Safety Standards for Independent 99 m Tc-MAA SPECT/CT Treatment Planning in Radioembolization
Significant improvements within radioembolization imaging and dosimetry permit the development of an accurate and personalized pretreatment plan using technetium 99m-labeled macroaggregated albumin ( Tc-MAA) and single-photon emission computed tomography (SPECT) combined with anatomical CT (SPECT/CT...
Gespeichert in:
Veröffentlicht in: | International journal of radiation oncology, biology, physics biology, physics, 2024-07, Vol.119 (4), p.1285 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Significant improvements within radioembolization imaging and dosimetry permit the development of an accurate and personalized pretreatment plan using technetium 99m-labeled macroaggregated albumin (
Tc-MAA) and single-photon emission computed tomography (SPECT) combined with anatomical CT (SPECT/CT). Despite these potential advantages, the clinical transition to pretreatment protocols with SPECT/CT is hindered by their unknown safety constraints. This study aimed to address this issue by establishing novel dose limits for
Tc-MAA SPECT/CT to enable quantitative pretreatment planning.
Stratification criteria to determine images most viable for dosimetry analysis were created from a cohort of 85 patients. SPECT/CT, cone beam CT, and activity calculations derived from the local deposition method were used to create an accurate pretreatment protocol. Planar and SPECT/CT images were compared using linear regression and modified Bland-Altman analyses to convert accepted planar dose limits to SPECT/CT. To validate these new dose limits, activity calculations based on SPECT/CT were compared with those calculated with the body surface area and planar methods for three treatment plans.
A total of 38 of 85 patients were deemed viable for dosimetry analysis. SPECT yielded greater lung shunt fractions (LSFs) than planar imaging when LSFs were 4.89%. Planar to SPECT/CT dose conversions were 0.76×, 0.70×, and 0.55× for the whole liver, normal liver, and lungs, respectively. Patients with SPECT LSFs ≤4.89% were safely treated with the direct application of planar lung dose limits. Activity calculations with the newly established SPECT/CT dose limits were greater than those of the body surface area method by a median range of 33.1% to 61.9% and were lower than planar-based activity calculations by a median range of 12.5% to 13.7% for the whole liver and by 29.4% to 32.2% for the normal liver.
This study demonstrated a safe method for translating dose limits from
Tc-MAA planar imaging to SPECT/CT. A robust pretreatment protocol was further developed guided by the current knowledge in the field. Established SPECT/CT dose limits safely treated 97.5% of patients and permitted the application of independent pretreatment planning with
Tc-MAA SPECT/CT. |
---|---|
ISSN: | 1879-355X |