Fluid-filled versus sensor-tipped pressure guidewires for FFR and P d /P a measurement; PW-COMPARE study
Fluid-filled pressure guidewires are unaffected by the previously inevitable hydrostatic pressure gradient (HPG). This study aimed to compare simultaneous pressure measurements with fluid-filled and sensor-tipped pressure guidewires. Fifty patients underwent fractional flow reserve (FFR) and P /P me...
Gespeichert in:
Veröffentlicht in: | International journal of cardiology 2024-07, Vol.406, p.131998 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Fluid-filled pressure guidewires are unaffected by the previously inevitable hydrostatic pressure gradient (HPG). This study aimed to compare simultaneous pressure measurements with fluid-filled and sensor-tipped pressure guidewires.
Fifty patients underwent fractional flow reserve (FFR) and P
/P
measurement with a fluid-filled and a sensor-tipped pressure guidewire simultaneously. To assess maneuverability, patients were randomized with respect to which pressure guidewire was used to cross the lesion first. Lateral fluoroscopy was used to estimate height difference between catheter tip and distal wire position (and thus HPG). Agreement between pressure measurements was studied.
Measurements were performed in LM (4% (n = 2)), LAD (44% (n = 22)), LCX (26% (n = 13)), and RCA (26% (n = 13)). Simultaneous pressure measurements showed excellent agreement (mean FFR difference - 0.01 ± 0.03 (r = 0.959, p |
---|---|
ISSN: | 1874-1754 |