Determination of 24,25-dihydroxyvitamin D 3 in Vitamin D External Quality Assessment Scheme samples using a reference measurement procedure
Ninety archived human serum samples from the Vitamin D External Quality Assessment Scheme (DEQAS) were analyzed using a reference measurement procedure (RMP) based on isotope dilution liquid chromatography - tandem mass spectrometry (ID LC-MS/MS) for the determination of 24,25-dihydroxyvitamin D [24...
Gespeichert in:
Veröffentlicht in: | The Journal of steroid biochemistry and molecular biology 2023-07, Vol.231, p.106318 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Ninety archived human serum samples from the Vitamin D External Quality Assessment Scheme (DEQAS) were analyzed using a reference measurement procedure (RMP) based on isotope dilution liquid chromatography - tandem mass spectrometry (ID LC-MS/MS) for the determination of 24,25-dihydroxyvitamin D
[24,25(OH)
D
]. These 24,25(OH)
D
results, in conjunction with concentration values assigned using RMPs for 25-hydroxyvitamin D
[25(OH)D
] and 25-hydroxyvitamin D
[25(OH)D
], provide a valuable resource for assessing the accuracy of measurements for 24,25(OH)
D
and for investigating the relationship between 24,25(OH)
D
and 25(OH)D
. Results for 24,25(OH)
D
using the RMP were compared to DEQAS consensus values demonstrating that the consensus values were not sufficient to assess the accuracy of measurements among different laboratories and methods. A multivariable regression analysis approach using historical DEQAS consensus values for various total 25(OH)D assays was used to assess the contribution of 24,25(OH)
D
concentration on the assay response. The response of several ligand binding assays for total 25(OH)D was shown to be impacted by the presence of 24,25(OH)
D
. |
---|---|
ISSN: | 1879-1220 |