A simple colorimetric method for viable bacteria detection based on cell counting Kit-8
In this study, Cell Counting Kit-8 (CCK-8) was introduced to detect the concentration of live bacteria for the first time depending on the redox reaction between CCK-8 solution and dehydrogenase. CCK-8 solution can be reduced to form water soluble orange-yellow formazan by the dehydrogenase present...
Gespeichert in:
Veröffentlicht in: | Analytical methods 2021-11, Vol.13 (43), p.5211-5215 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | In this study, Cell Counting Kit-8 (CCK-8) was introduced to detect the concentration of live bacteria for the first time depending on the redox reaction between CCK-8 solution and dehydrogenase. CCK-8 solution can be reduced to form water soluble orange-yellow formazan by the dehydrogenase present in bacterial cells, and the concentration of live bacteria is proportional to the absorbance value of formazan at 450 nm. Based on this principle,
Staphylococcus aureus
and
Escherichia coli
were chosen as the model bacteria. The optimal detection conditions were investigated and a good linear relationship was obtained in the concentration range from 2.600 × 10
2
to 1.160 × 10
9
CFU mL
−1
with a linear equation of
Y
= 0.06305 log
10
X
-0.1153 (
X
in CFU mL
−1
,
R
2
= 0.9747) for
S. aureus
and 9.750 × 10
2
to 6.000 × 10
8
CFU mL
−1
with a linear equation of
Y
= 0.06122 log
10
X
-0.1358 (
X
in CFU mL
−1
,
R
2
= 0.9958) for
E. coli
. The CCK-8 based viable bacteria detection method can be completed within 2 h with a wide bacterial detection concentration range. Satisfactory results were obtained when applied to an actual sample analysis and there is a good consistency between the proposed CCK-8 based method and the traditional plate counting method. More importantly, this method can realize the one-time detection of a large number of samples with high sensitivity, which suggests its great potential in high-throughput bacterial detection.
Cell counting Kit-8 was used to detect viable bacteria based on the redox reaction between the CCK-8 solution and dehydrogenase in bacteria. The method can detect a large number of samples at a time, with high sensitivity, good reliability and repeatability. |
---|---|
ISSN: | 1759-9660 1759-9679 |
DOI: | 10.1039/d1ay01624e |