WRISTOX 2 is a reliable tool to diagnose obstructive sleep apnoea syndrome
Overnight pulse oximetry is an alternative to polysomnography (PSG) in diagnosing obstructive sleep apnoea syndrome, but its sensitivity is reported to be low. To determine the (a) diagnostic accuracy, interobserver reliability and reliable oxygen desaturation index of 4% (ODI ) score at diagnosing...
Gespeichert in:
Veröffentlicht in: | International journal of pediatric otorhinolaryngology 2021-12, Vol.151, p.110930 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Overnight pulse oximetry is an alternative to polysomnography (PSG) in diagnosing obstructive sleep apnoea syndrome, but its sensitivity is reported to be low.
To determine the (a) diagnostic accuracy, interobserver reliability and reliable oxygen desaturation index of 4% (ODI
) score at diagnosing obstructive sleep apnoea syndrome in children and (b) correlation between the apnoea hypopnoea index (AHI) with ODI
and oxygen nadir between both PSG and oximetry.
This cross-sectional study included children aged 1-18 years old, undergoing a fully attended overnight PSG for suspected obstructive sleep apnoea syndrome. The Nonin 3150 WristOx
™ [Fig. 2] was worn simultaneously during the PSG. Poor oximetry recordings were excluded. Pulse oximetry was scored using the McGill Oximetry Score (MOS) whereby a score of 2-4 was positive for OSAS. Specificity, sensitivity, positive predictive values (PPV), negative predictive values (NPV) and interobserver reliability of the WristOx
were calculated.
One hundred and sixty-two children with a mean (SD) age of 9.3 (±3.5) years (range 2 years 6 months old - 17 years old) were included after excluding 18 children (poor oximetry data [n = 16] and incomplete PSG [n = 2]). Interobserver agreement of the WristOx
was 0.8763 (95% CI:0.80, 0.95). WristOx
had a sensitivity 50%, specificity 96.7%, PPV 96% and NPV 53% at diagnosing OSAS. ODI
≥ 2 events/hour in oximetry had a sensitivity of 97.6% and negative predictive value of 85.7% at diagnosing OSA.
Overnight pulse oximetry with the Nonin 3150 WristOx
™ is an accurate and reliable tool in diagnosing significant OSAS in children. |
---|---|
ISSN: | 1872-8464 |
DOI: | 10.1016/j.ijporl.2021.110930 |