The OralChroma TM CHM-2: a comparison with the OralChroma TM CHM-1
Already decades ago, the diagnosis of halitosis was facilitated with the arrival of chair-side instruments to score the breath odor. These devices are used for a more objective assessment of halitosis compared with organoleptic scoring, but these too have their disadvantages. To overcome some of the...
Gespeichert in:
Veröffentlicht in: | Clinical oral investigations 2020-01 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Already decades ago, the diagnosis of halitosis was facilitated with the arrival of chair-side instruments to score the breath odor. These devices are used for a more objective assessment of halitosis compared with organoleptic scoring, but these too have their disadvantages. To overcome some of the drawbacks of the original model of the OralChroma
(CHM-1), few years ago a second generation of this machine (CHM-2) was introduced. This study compared both devices in a clinical setting.
All records of the patients visiting a specialized halitosis consultation over a period of 5 years (2012-2017) were examined. The correlations of the OralChroma
CHM-1 and CHM-2 with the organoleptic and Halimeter® measurements were analyzed. Additionally the sensitivity, specificity, and negative and positive predicted values were calculated.
A total of 581 data points were included (CHM-1: 292, CHM-2: 289). The correlations between both models with the organoleptic measurements were not statistically significant different. The CHM-2 seemed superior to the CHM-1 in the quantification of dimethyl sulfide with a detection rate of 95% and 61%, respectively. Additionally, the CHM-2 was significantly more sensitive for dimethyl sulfide than the CHM-1. However the latter showed in turn a better sensitivity for hydrogen sulfide and methyl mercaptan.
The CHM-2 showed a better sensitivity for dimethyl sulfide than its predecessor. However, its sensitivity for hydrogen sulfide and methyl mercaptan was worse.
Dimethyl sulfide is the main volatile implicated in extra-oral blood-borne halitosis, this makes the OralChroma
CHM-2 the instrument of choice when this is suspected. |
---|---|