Mini EXPLORER II: a prototype high-sensitivity PET/CT scanner for companion animal whole body and human brain scanning

As part of the EXPLORER total-body positron emission tomography (PET) project, we have designed and built a high-resolution, high-sensitivity PET/CT scanner, which is expected to have excellent performance for companion animal whole body and human brain imaging. The PET component has a ring diameter...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Physics in medicine & biology 2019-03, Vol.64 (7), p.075004-075004
Hauptverfasser: Lv, Yang, Lv, Xinyu, Liu, Weiping, Judenhofer, Martin S, Zwingenberger, Allison, Wisner, Erik, Berg, Eric, McKenney, Sarah, Leung, Edwin, Spencer, Benjamin A, Cherry, Simon R, Badawi, Ramsey D
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:As part of the EXPLORER total-body positron emission tomography (PET) project, we have designed and built a high-resolution, high-sensitivity PET/CT scanner, which is expected to have excellent performance for companion animal whole body and human brain imaging. The PET component has a ring diameter of 52 cm and an axial field of view of 48.3 cm. The detector modules are composed of arrays of lutetium (yttrium) oxyorthosilicate (LYSO) crystals of dimensions 2.76  ×  2.76  ×  18.1 mm3 coupled to silicon photomultipliers (SiPMs) for read-out. The CT component is a 24 detector row CT scanner with a 50 kW x-ray tube. PET system time-of-flight resolution was measured to be 409  ±  39 ps and average system energy resolution was 11.7%  ±  1.5% at 511 keV. The NEMA NU2-2012 system sensitivity was found to be 52-54 kcps MBq−1. Spatial resolution was 2.6 mm at 10 mm from the center of the FOV and 2.0 mm rods were clearly resolved on a mini-Derenzo phantom. Peak noise-equivalent count (NEC) rate, using the NEMA NU 2-2012 phantom, was measured to be 314 kcps at 9.2 kBq cc−1. The CT scanner passed the technical components of the American College of Radiology (ACR) accreditation tests. We have also performed scans of a Hoffman brain phantom and we show images from the first canine patient imaged on this device.
ISSN:0031-9155
1361-6560
1361-6560
DOI:10.1088/1361-6560/aafc6c