In vivo Evaluation of Physiological Control Algorithms for LVADs based on Left Ventricular Volume or Pressure

Turbodynamic left ventricular assist devices (LVADs) provide a continuous flow depending on the speed at which the pump is set, and do not adapt to the changing requirements of the patient. The limited adaptation of the pump flow to the amount of venous return can lead to ventricular suction or over...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:ASAIO journal (1992) 2017-02
Hauptverfasser: Ochsner, Gregor, Wilhelm, Markus J, Amacher, Raffael, Petrou, Anastasios, Cesarovic, Nikola, Staufert, Silvan, Röhrnbauer, Barbara, Maisano, Francesco, Hierold, Christofer, Meboldt, Mirko, Daners, Marianne Schmid
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Turbodynamic left ventricular assist devices (LVADs) provide a continuous flow depending on the speed at which the pump is set, and do not adapt to the changing requirements of the patient. The limited adaptation of the pump flow to the amount of venous return can lead to ventricular suction or overload. Physiological control may compensate such situations by an automatic adaptation of the pump flow to the volume status of the left ventricle. We evaluated two physiological control algorithms in an acute study with eight healthy pigs. Both controllers imitate the Frank-Starling law of the heart and are based on a measurement of the left ventricular volume or pressure, respectively. After implantation of a modified Deltastream DP2 blood pump as an LVAD, we tested the responses of the physiological controllers to hemodynamic changes and compared them with the response of the constant speed mode. Both physiological controllers adapted the pump speed such that the flow was more sensitive to preload and less sensitive to afterload, as compared to the constant speed mode. As a result, the risk for suction was strongly reduced. Five suction events were observed in the constant speed mode, one with the volume-based controller, and none with the pressure-based controller. The results suggest that both physiological controllers have the potential to reduce the number of adverse events when used in the clinical setting.
ISSN:1058-2916
1538-943X
DOI:10.1097/MAT.0000000000000533