Aptamer-based fiber sensor for thrombin detection

The detection of thrombin based on aptamer binding is studied using two different optical fiber-based configurations: long period gratings coated with a thin layer of titanium dioxide and surface plasmon resonance devices in optical fibers coated with a multilayer of gold and titanium dioxide. These...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of biomedical optics 2016-08, Vol.21 (8), p.087005-087005
Hauptverfasser: Coelho, Luís, Marques Martins de Almeida, José Manuel, Santos, José Luís, da Silva Jorge, Pedro Alberto, Martins, Maria Cristina L, Viegas, Diana, Queirós, Raquel B
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The detection of thrombin based on aptamer binding is studied using two different optical fiber-based configurations: long period gratings coated with a thin layer of titanium dioxide and surface plasmon resonance devices in optical fibers coated with a multilayer of gold and titanium dioxide. These structures are functionalized and the performance to detect thrombin in the range 10 to 100 nM is compared in transmission mode. The sensitivity to the surrounding refractive index (RI) of the plasmonic device is higher than 3100  nm RIU−1 in the RI range 1.335 to 1.355, a factor of 20 greater than the sensitivity of the coated grating. The detection of 10 nM of thrombin was accomplished with a wavelength shift of 3.5 nm and a resolution of 0.54 nM.
ISSN:1083-3668
1560-2281
DOI:10.1117/1.JBO.21.8.087005