Performance analysis of a dedicated breast MR-HIFU system for tumor ablation in breast cancer patients

MR-guided HIFU ablation is a promising technique for the non-invasive treatment of breast cancer. A phase I study was performed to assess the safety and treatment accuracy and precision of MR-HIFU ablation in breast cancer patients () using a newly developed MR-HIFU platform dedicated to application...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Physics in medicine & biology 2015-07, Vol.60 (14), p.5527-5542
Hauptverfasser: Deckers, R, Merckel, L G, Denis de Senneville, B, Schubert, G, Köhler, M, Knuttel, F M, Mali, W P Th M, Moonen, C T W, van den Bosch, M A A J, Bartels, L W
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:MR-guided HIFU ablation is a promising technique for the non-invasive treatment of breast cancer. A phase I study was performed to assess the safety and treatment accuracy and precision of MR-HIFU ablation in breast cancer patients () using a newly developed MR-HIFU platform dedicated to applications in the breast. In this paper a technical analysis of the performance of the dedicated breast MR-HIFU system during breast tumors ablation is described. The main points of investigation were the spatial targeting accuracy and precision of the system and the performance of real-time respiration-corrected MR thermometry. The mean targeting accuracy was in the range of 2.4-2.6 mm, whereas the mean targeting precision was in the range of 1.5-1.8 mm. To correct for respiration-induced magnetic field fluctuations during MR temperature mapping a look-up-table (LUT)-based correction method was used. An optimized procedural sedation protocol in combination with the LUT-based correction method allowed for precise MR thermometry during the ablation procedure (temperature standard deviation
ISSN:0031-9155
1361-6560
DOI:10.1088/0031-9155/60/14/5527