A slit method to determine the focal spot size and shape of TomoTherapy system
Purpose : To obtain accurate x-ray source profile measurements using a slit-collimator, the slit-collimator should have a narrow width, large height, and be positioned near the source. However, these conditions may not always be met. In this paper, the authors provide a detailed analysis of the slit...
Gespeichert in:
Veröffentlicht in: | Medical physics (Lancaster) 2011-06, Vol.38 (6), p.2841-2849 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Purpose
: To obtain accurate x-ray source profile measurements using a slit-collimator, the slit-collimator should have a narrow width, large height, and be positioned near the source. However, these conditions may not always be met. In this paper, the authors provide a detailed analysis of the slit measurement geometry and the relationship between the slit parameters and the measured x-ray source profile. The slit model allows the use of a shorter and more easily available slit-collimator, while accurate source profile measurements can still be obtained.
Methods
: Measurements were performed with a variety of slit widths and/or slit to source distances. The relationship derived between the slit parameters and the measured profile was used to determine the true focal spot profile through a least square fit of the profile data. The model was verified by comparing the predicted profiles at a variety of slit-collimator parameters with the measured results on the TomoTherapy Hi-Art system.
Results
: Both the treatment beam and the imaging beam were measured. For treatment mode, it was found that a source consisting of one Gaussian with a 0.75 mm full-width-half-maximum (FWHM) and 72% peak amplitude and a second Gaussian with a 2.27 mm FWHM and 18% peak amplitude matched measurement profiles. The overall source profile has a FWHM of 0.93 mm, but with a higher amplitude in the tail region than a single Gaussian. For imaging mode, the source consists of one Gaussian with a 0.68 mm FWHM and 82% peak amplitude and a second Gaussian with a 1.83 mm FWHM and 18% peak amplitude. The overall source profile has a FWHM of 0.77 mm.
Conclusions
: Our study of the focal spot measurement using slit-collimators showed that accurate source profile measurements can be achieved through fitting of measurement results at different slit widths and source-to-slit distances (SSD). Quantitative measurements of the TomoTherapy linac focal spot showed that the source distribution could be better described with a model consisting of two Gaussian components rather than a single Gaussian model as assumed in previous studies. |
---|---|