Aliphatic polyester terpolymers for stent coating and drug elution: Effect of polymer composition on drug solubility and release
Ring-opening terpolymerization of L-lactide (LA), -caprolactone (CL), and glycolide (GA) was performed in the presence of tin (II) 2-ethylhexanoate at 170°C. Random terpolyesters with weight-average molecular weight up to 130,000 g/mol were obtained. These terpolyesters, especially those with LA:CL:...
Gespeichert in:
Veröffentlicht in: | Drug delivery 2009-08, Vol.16 (6), p.304-311 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Ring-opening terpolymerization of L-lactide (LA), -caprolactone (CL), and glycolide (GA) was performed in the presence of tin (II) 2-ethylhexanoate at 170°C. Random terpolyesters with weight-average molecular weight up to 130,000 g/mol were obtained. These terpolyesters, especially those with LA:CL:GA composition of 3:1:1, provided good coating integrity following spraying onto bare metal stents. The semi-synthetic macrolide immunosuppressant, everolimus, was incorporated into the terpolyester coating, and its release from the stent was evaluated. Unlike PLLA homopolymers, which are immiscible with the drug and non-optimal for controlled release, these terpolymers gave excellent control in a screening study, by tuning terpolymer molecular weight, relative monomer ratio, and drug-to-polymer ratio. Adjusting the polymer properties to improve drug solubility (or miscibility) in the polymer coating was found beneficial to the release profile. |
---|---|
ISSN: | 1071-7544 1521-0464 |
DOI: | 10.1080/10717540902999653 |