A Comparison of Analytical Goals for Haemoglobin A1c Assays Derived Using Different Strategies
Analytical goals for the performance characteristics of assays of haemaoglobin A1c (HbA1c) have been investigated using different assumptions for generation of estimates, these being based on strategies using data on biological variation and on the clinical use of results. The derived goals are high...
Gespeichert in:
Veröffentlicht in: | Annals of clinical biochemistry 1991-05, Vol.28 (3), p.272-278 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Analytical goals for the performance characteristics of assays of haemaoglobin A1c (HbA1c) have been investigated using different assumptions for generation of estimates, these being based on strategies using data on biological variation and on the clinical use of results. The derived goals are highly dependent on the assumptions made. In general, in monitoring of patients (using results from the same laboratory), the analytical imprecision is the most demanding, whereas bias (inaccuracy) is the most important characteristic when strategies for several centres (laboratories) to achieve similar results are invoked. Goals for analytical quality should be given in a form in which both analytical imprecision and bias (or systematic error) are specified. When several goals are to be considered (for different relevant assumptions), the most demanding should be used. |
---|---|
ISSN: | 0004-5632 1758-1001 |
DOI: | 10.1177/000456329102800313 |